-
1.
Remove the clear acrylic bar top from the cardboard
envelope and place it on top of the bar. There is a
small notch on the wide side of the bartop, align this
with the center line of the back doors. Be sure the back
of the bar and the back of the bartop are flush. The Velcro
strips will hold the top. It is best not to lift the bar
with the top as the Velcro may release and the bar could
be dropped.
-
-
2.
(we
now do this step for you at the factory)
in your starter kit is the connector kit. Connect the
line from the center "Tee" to the oxygen
Generators Output barb. Zip tie the connector kit along
the light bar immediately above. Be sure that the tubing
can reach about 6 inches through each hole in the bartop.
Use a line connector and cut extra lengths of tubing
from the 25' bulk tubing in the starter kit if necessary.
-
-
3.
(we
now do this step for you at the factory)
Locate the 4 leads with the 2.5mm X 5.5 mm jacks on the
end. It looks nicer if you also string these with the
vinyl tubing along the light bar. They need to stick out
a little longer than the tubing does as they connect to
the back of the infusor and the vinyl tubing connects to
the bottom a 4 inch shorter distance.
-
-
4.
Place the four Infusors on the bartop and connect them
to the power jacks and the vinyl supply tubing as shown.
It is better if you put the filled aroma bottles in
after you connect as you need to lean the infusor over
to connect the oxygen supply tube and you could dump the
aroma.
-
-
5.
Connect power to the bar. A 2" hole is provided in
the bottom of the bar for the power cord to go out to an extension cord or wall plug if close enough. Turn on
the power strip. The fans and light should come on. We
ship the bar with the Generator switch in the off
position, if it is on,
turn it off now. We want to check that the infusors are
powering properly before running oxygen through them.
-
-
6.
Check that the baselite in each infusor is on and that
the digital timer scrolls "O2 Bar" press the
green button on each infusor a couple of times and check
that time is counting up. Wait for 2 seconds and you
should hear a soft click come from each unit. That is
the "oxygen on" solenoid activating.
-
-
7.
Fill your aroma bottles with the aromas of your choice
DO NOT put the aroma labels on them until you have
installed them in the Infusors or the labels will not
face the customer. Install the aromas in the
infusors by aligning the bottles under the holes in the
upper manifold and twisting to the right while putting
up pressure. Once the bottles start to thread turn them
until they are snug. Put some time on the timer and turn on one or
more of the aromas on the gang valve on the front. If
everything checks out cut a length of tubing about
24" long to use as a supply tube push it on to the
chrome output barb on the front. Use this to connect the
Oxerator the customer puts on.
-
-
8.
Do a full load test now. Put a full 30 minutes on each
Infusor and turn on all 4 aromas on each turn on the
generator and let each Infusor run to zero.